| Catalog Number |
309.530 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E1: initial reporter state: new south wales e3: reporter is a j&j employee.H3, h4, h6: the product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that the device 309.530, extract-screw coni f/scr ø4.5+6.5 had broken tip.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to excessive/unintended forces ¿ as the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the device 309.530, extract-screw coni f/scr ø4.5+6.5 would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): product code: 309.530.Lot number : l745846.Release to warehouse date : 05.Mar.2018.Expiration date : na supplier: na manufacturing site: werk bettlach a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in australia as follows: it was reported that on (b)(6) 2023, it was identified that the tips of the instruments had been snapped off.This was noted by the loan kit technician during loan kit inspection.No further details are available.This report involves one conical extraction screw for large screws & 4.9mm bolts.This is report 1 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned device found that the extract-screw coni f/scr ø4.5+6.5 was broken from the distal area.The broken fragment was not returned for evaluation.No other issues were observed the observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.A dimensional inspection for the extract-screw coni f/scr ø4.5+6.5 was unable to be performed due to post manufacturing damage.The overall complaint was confirmed as the observed condition of the extract-screw coni f/scr ø4.5+6.5 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Device history review (dhr): part # 309.530, lot # l745846, manufacturing site: werk bettlach, release to warehouse date : 03 march 2018, expiration date: n/a, supplier: n/a.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|