| Catalog Number |
08056722190 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The analyzer serial number is (b)(6).The field service engineer (fse) replaced the sample probe, checked the cell wash station, cleaned the flow path, readjusted a probe, and verified rinsing volumes for the measuring cells.The service maintenance actions performed by the fse resolved the issue.No further issues were reported after the service visit.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable albt2 albumin gen.2 results for 2 patient samples on a cobas pro c 503 analytical unit.The customer was prompted to repeat the samples as the results were not consistent with the patients' clinical pictures.On (b)(6) 2023, sample 1 had an initial result of 3.77 g/l.The repeat result was 39.9 g/l.On (b)(6) 2023, sample 2 had an initial result of <3 g/l with flag.The sample was repeated twice and the results were 20 g/l and 20 g/l.No erroneous results were reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The medical coding in this case has been updated.
|
| |
|