MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON IOL WITH THE AUTONOME DELIVERY SYSTEM; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON IOL WITH THE AUTONOME DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number CNA0T0
Medical Device Problem Code Particulates (1451)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
A healthcare professional reported that, during intraocular lens implantation surgery, the physician white sheen on the anterior surface of the lens.The white sheen remains even after the removal of the ophthalmic viscoelastic device.The lens was left in the patients eye and will be monitored.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The product was not returned for analysis.One video provided for multiple records to show the reported complaint.Photo & video review: white in appearance shadow is seen over implanted iol.Based on our observation of the provided video, white in appearance shadow is seen over implanted iol.No root cause identified to date.No reports of impact to visual function in any of the complaints reported.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON IOL WITH THE AUTONOME DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17870385
Report Number9612169-2023-00694
Device Sequence Number12184066
Product Code HQL
UDI-Device Identifier00380652393652
UDI-Public00380652393652
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/26/2026
Device Model NumberCNA0T0
Device Lot Number25618993
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/08/2023
Supplement Date Received by Manufacturer01/04/2024
Initial Report FDA Received Date10/04/2023
Supplement Report FDA Received Date01/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/28/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PROVISC OPHTHALMIC VISCOSURGICAL DEVICE.
Patient SexUnknown
-
-