• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ORMCO CORPORATION SPARK ADVANCED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ORMCO CORPORATION SPARK ADVANCED Back to Search Results
Catalog Number 728-3031
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Abscess (1690)
Date of Event 09/07/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a patient had pain in the ll anterior region and also had migraines.Upon examination the doctor diagnosed an abscess in the #22 area with significant facial movement in that area and significant bone loss.Doctor recommended to patient to stay in current aligners and not move forward until infection clears and doctor speak with spark re about the movements for this patient.Patient experimented the following symptoms: pain in #22 area, swelling, irritation, severe headaches, tooth mobility and severe bone loss.In doctor opinion, this complaint would it likely cause a serious injury because the patient would be prone to another infection and suffer permanent tooth loss in #22 or even all of her lower anterior teeth.Patient started spark aligners on (b)(6) 2022.Her first refinement was delivered on (b)(6) 2023 and she was given aligners stages 1 to 7 of 1 week.Patient came back on (b)(6) 2023 and was given aligners stages 8 to 14 and called on (b)(6) 2023 and was on aligner stage 14 at that point.It was prescribed clindamycin and srp with minocycline placement.Patient is not fully recovered because pain and headaches have subsided but doctor is extremely concerned with her bone loss and possible loss of permanent teeth.
 
Additional Manufacturer Narrative
Based on the x-rays provided by the doctor, the bone level is the same as the patient already had general bone resorption before starting the treatment plan and general root resorption.Based on evidence, the patient had a pre-existing condition in tooth 22, root resorption with radiolucency in the apex of 22 (endo- and/or periodontal involvement).Due to the patient's bone level at the beginning of treatment, aligners should not have been used because there is a warning in aligner ifu pn 070-010 rev.H mentions: the health of the bone and gums which support the teeth may be impaired or aggravated.The lenghth of the roots of the teeth may be shortened during orthodontic treatment, which may become a threat to the longevity of the teeth.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPARK ADVANCED
Common Device Name
SPARK ADVANCED
Manufacturer (Section D)
ORMCO CORPORATION
200 s kraemer blvd
brea CA 92821
Manufacturer (Section G)
SDS DE MEXICO
cto. sur 31, nelson
mexicali, baja california 21395
MX   21395
Manufacturer Contact
procoro herrera
200 s kraemer blvd
brea, CA 92821
7148174396
MDR Report Key17870867
Report Number2016150-2023-00006
Device Sequence Number7028340
Product Code NXC
UDI-Device Identifier00889989084856
UDI-Public00889989084856
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 09/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue Number728-3031
Device Lot Number746266
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/13/2023
Initial Report FDA Received Date10/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age57 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-