| Catalog Number |
CAT02298 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke during the procedure.The piece was retrieved.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke during the procedure.The piece was retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacture date is not known.Alleged failure: piece of jaw from the nanopass broke off during labral repair.The failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be: user technique related factors, difficult anatomy, extremely tough soft tissue, use of excessive force.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|