|
At the time of this report, mentor has received no information regarding explantation or an expected explantation date. it is unknown at this time if the device will be made available for return. as a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made. as such, the investigation will be closed. if the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803. this report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date. this report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report. if certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: the patient has not undergone explantation or reoperation at this time.Manufacturer¿s reference number: (b)(4).
|
|
It was reported that a patient underwent a breast augmentation revision surgery with implantation of a 275/300cc mentor siltex contour profile moderate saline breast implant prosthesis.Post-operatively, the patient reported her healthcare professional requested a mammogram to be performed due to bilateral breast cysts.After the mammogram was performed, her breasts felt different and became painful.Additionally, she reports bilaterally deflated breast implants.Magnetic resonance imaging and a consultation with a medical professional have been conducted; however, at the time of this report, mentor has received no information regarding explantation or an expected explantation date.Refer to manufacturing report number 1645337-2023-11794 for the contralateral event.
|