• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S FOLYSIL SILICONE CATHETER; INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S FOLYSIL SILICONE CATHETER; INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL Back to Search Results
Catalog Number AA6418
Medical Device Problem Codes Deflation Problem (1149); Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Event or Problem Description
According to available information, this device was difficult to inflate and deflate.During the inflation test before device placement, it was difficult to inject and remove water from the balloon.No other adverse patient effects were reported.
 
Additional Manufacturer Narrative
After receiving this complaint, we searched for other complaint and we didn't find other complaint regarding the lot number 9018807.In september, we received one used sample.After desinfection it was observed that the balloon was assymetric and blocked the way to inflate/deflate and caused the issue observed.Our subcontractor investigation concluded on raw material issue on balloons conducting to assymetry.Checking the quality databases revealed one corrective and preventive action are ongoing: monitoring capa-000152: "balloon issues on folysil and silicone prostatic catheters".The trending on balloon asymetry issue is followed and below the threshold.
 
Event or Problem Description
According to available information, this device was difficult to inflate and deflate.During the inflation test before device placement, it was difficult to inject and remove water from the balloon.No other adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLYSIL SILICONE CATHETER
Common Device Name
INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer (Section G)
CMF-SARLAT
9 avenue edmond rostand
sarlat-la-caneda
FR  
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key17874035
Report Number9610711-2023-00207
Device Sequence Number4320947
Product Code EZL
UDI-Device Identifier03600040130087
UDI-Public3600040130087
Combination Product (Y/N)N
PMA/510(K) Number
K013174
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberAA6418
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/04/2023
Supplement Report FDA Received Date04/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-