| Model Number |
MA60AC |
| Medical Device Problem Codes |
Defective Component (2292); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Headache (1880); Eye Pain (4467)
|
| Date of Event |
08/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was returned for analysis.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an intraocular lens (iol) implant procedure, findings on both follow up visits showed well centered iol, clear cornea & corneal suture in place on operated eye.The patient came with complain of head ache and pain on the operated eye.After the examination, the surgeon found out that the haptic slipped from the junction on optic.Additional information has been requested, received and stated explant was planned.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned.Product history records were reviewed and documentation indicated the product met release criteria.A non-company viscoelastic was indicated.A cartridge and handpiece were not indicated.It is unknown if a delivery system was used or if the lens was manually implanted.The product investigation could not identify a root cause for the reported complaint.A determination cannot be made without physical examination of the product.The company 30.0 diopter lens is qualified for use with the company cartridges.The instruction for use (ifu) instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The reporter has not responded to the follow up requests.The file will be reopened if the sample or new information is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|