ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF MULTIPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number MA60AC
Medical Device Problem Codes Defective Component (2292); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Headache (1880); Eye Pain (4467)
Date of Event 08/30/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was returned for analysis.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A facility representative reported that following an intraocular lens (iol) implant procedure, findings on both follow up visits showed well centered iol, clear cornea & corneal suture in place on operated eye.The patient came with complain of head ache and pain on the operated eye.After the examination, the surgeon found out that the haptic slipped from the junction on optic.Additional information has been requested, received and stated explant was planned.
 
Additional Manufacturer Narrative
The product was not returned.Product history records were reviewed and documentation indicated the product met release criteria.A non-company viscoelastic was indicated.A cartridge and handpiece were not indicated.It is unknown if a delivery system was used or if the lens was manually implanted.The product investigation could not identify a root cause for the reported complaint.A determination cannot be made without physical examination of the product.The company 30.0 diopter lens is qualified for use with the company cartridges.The instruction for use (ifu) instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The reporter has not responded to the follow up requests.The file will be reopened if the sample or new information is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF MULTIPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17874832
Report Number1119421-2023-01708
Device Sequence Number4386333
Product Code HQL
UDI-Device Identifier00380652552721
UDI-Public00380652552721
Combination Product (Y/N)N
Initial Reporter CountrySA
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMA60AC
Device Catalogue NumberMA60AC.300
Device Lot Number15505334
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/12/2023
Supplement Date Received by Manufacturer10/17/2023
Initial Report FDA Received Date10/05/2023
Supplement Report FDA Received Date11/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/03/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OPHLINE 2%
Patient SexUnknown
-
-