| Catalog Number |
8065751614 |
| Medical Device Problem Code |
Suction Failure (4039)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A health care professional reported that the vitrector stopped working and not sucking during surgery.No patient harm was reported.Additional information was requested and received that the surgery was completed after replacing the product with another one.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
A sample was not received at the manufacturing site for evaluation.Therefore, the condition of the product could not be verified.A review of the device history records traceable to the reported lot number indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A sample was not received at the manufacturing site and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown; therefore, specific action with regards to this complaint cannot be taken.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
One opened probe was received with a tip protector in a bag.The sample was visually inspected and found to be conforming.The sample was then functionally tested for actuation and aspiration.The sample was found to be conforming for both functional tests.The probe was disassembled, and the components inspected.No/minimal wear observed on inner cutter when compared to the degree of wear based on continuous actuation of the probe visual standard photos.Gouge marks observed at several locations along the inner cutter.The sample was retested for actuation with the probe driver and was able to actuate.The initial actuation test failed due to an interference within the probe and once the interference (needle assembly) was removed the probe was able to actuate.The needle holder and retainer were dissembled and visually inspected, all components were deemed visually conforming.A review of the device history record traceable to the lot number obtained from the device¿s radio frequency identification (rfid) tag, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A photo attached was reviewed by the investigation site.The photo shows a pak label with same product and lot number as reported.The complaint evaluation does not confirm the probe had an aspiration failure, the evaluation indicates the probe had an actuation failure.The aspiration function of the probe was conforming, however, the observed actuation failure could have caused the cutter to stay in the port of the needle long enough to obstruct aspiration flow at the time the reported event was observed.The exact cause of the actuation failure cannot be determined from the evaluation performed.The reported aspiration failure was not confirmed, and an exact root cause for the actuation failure could not be determined from this evaluation; therefore, no specific action with regard to this complaint was taken by the manufacturing site.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Any non-conformances found are removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: 2023-62573.
|
| |
|