This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4, b5, d2, g1, g3, g6, h1, h2, h3, h6, h10 h6-component code- suggested code: mechanical (g04) - provisional bottom visually verified item lot combination and reviewed manufacturing date as instrument exhibits signs of repeated use (nicks, gouges) and anterior section of the post feature is fractured, all pieces returned.The device history records & receiving inspection report were reviewed for deviations and / or anomalies with no deviations / anomalies identified.Reported event is not related to a combination of products; therefore, a compatibility review is not applicable.Medical records were not provided.Fractured tasps were analyzed by optical microscopy revealing hackle marks and river lines in the case of bending overload; and hackle marks, river lines and striations were found in the case of low cycle fatigue culminating in a bending overload failure type.Complaint is confirmed via product return & evaluation.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|