| Catalog Number |
NSEAL535RH |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a sigmoid colectomy procedure that the generator did not give a sound when the device was done sealing.Another like device was used to continue with the procedure.There were no adverse consequences for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 10/5/2023.D4 batch #: a9ce1k.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was received with the top of the i-blade fractured and slightly lifted.The device was connected to the generator and it was recognized.Because the i blade was damaged not all functional testing could be performed with the generator.The jaws were found attached to the instrument.In addition no pieces were found missing under microscope inspection.The condition of the blade prevented the functionality of the jaw.The jaw was unable to cycle open and close.A probable cause of the damage to the i blade could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.As described in the ifu: do not use excessive force on the closing handle to close the jaws; grasp only as much tissue as will fit between the jaws where the current will pass.Greater amounts of tissue require more closing handle force.Excessive force could damage the device.As part of the quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.A manufacturing record evaluation was performed for the finished device batch number a9ce1k, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|