The customer reported to olympus, at the outpatient surgery unit, during procedure, it was reported that the device-ureteroscope, 6,4/7,8 fr.X 330 mm, 7°, straight ocular angle, 4,2 fr.Channel, with wa00395a was bent at the end of the procedure when removing the scope.The procedure was ureteroscopy laser lithotripsy stenting undertaken and completed with the same device.The procedure was prolonged by 15 mins.Though there was a delay, this delay was reported to be a short one.There were no reports of patient harm or clinical impact associated with the reported event.No medical intervention was undertaken to preclude permanent harm/injury to the patient.There is no evidence of any serious deterioration in the state of health of the patient.Also, no indication of any life-threatening event occurred during the procedure.
|
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.New information added to the following fields: a review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over one (1) year since the subject device was manufactured.The device was returned, and an evaluation was completed for it.During inspection, olympus confirmed the reported event.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely the event occurred due excessive use.Olympus will continue to monitor field performance for this device.
|