| Catalog Number |
07528582190 |
| Medical Device Problem Codes |
High Test Results (2457); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The analyzer serial number was (b)(6).The quality control results were acceptable.Therefore a general reagent problem could be ruled out.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable hdlc4 results from the cobas 8000 c702 module.The first result was 113.7 mg/dl and the repeat results on another system were 37.7 mg/dl and 37.8 mg/dl.The questionable result was not reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
Calibration was last performed on 13-sep-2023 with acceptable results.A general reagent issue can be excluded as calibration and qc were acceptable.The alarm trace showed frequent abnormal aspiration alarms.Based on the frequent abnormal aspiration alarms, the root cause of the event was found to be consistent with pre-analytical sample handling issues.
|
| |
|