| Catalog Number |
MESE1 |
| Medical Device Problem Code |
Difficult or Delayed Activation (2577)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 10/5/2023.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Serial number was received and dhr is pending review.When the review is completed, a supplemental medwatch will be sent with a summary of the evaluation.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure the mese1 was malfunction.When i power on the unit without rf sensor and try to test the functional by pressed the manual button mese1 was function normally.But if i power on the unit with rf sensor together with megen1 and telescoping smoke evacuation soft tissue dissector; i try to pressed the dissector by handswitch mese1 doesn?t function.If i power on the unit and connect rf sensor at the back of the mese1 the mese1 will function automatically without pressing the dissector.Serial number : (b)(6).There were no patient status/ outcome / consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 11/29/2023.Investigation summary: per service manual operational and diagnostic analysis confirmed the activation and self activation issues.The pcb was replaced to address the activation issue and the rear panel foot control sensor was replaced to address the self activation as identified in the investigation to address the issue.The repair and testing of the unit was completed per the service manual, bringing the unit back to full functionality.The unit passed all functional tests and is fully operational.The repair and testing of the unit was completed.The device history records were reviewed and certed by external manufacturing that the manufacturing criteria were met prior to the release of the equipment.The certificate records are accessible through external manufacturing.
|
| |
|
Search Alerts/Recalls
|