MAUDE Adverse Event Report: ATRICURE, INC. ENCOMPASS SYNERGY CLAMP (LONG)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ATRICURE, INC. ENCOMPASS SYNERGY CLAMP (LONG) Back to Search Results
Model Number OLH
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2023
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2023, a patient underwent an aortic valve, myectomy, and concomitant maze procedure.During the maze portion of the procedure, the blue plunger insert of the closure lever detached when the clamp was released and fell into the sterile field.The procedure was completed with the device.There was no reported patient injury.
 
Additional Manufacturer Narrative
(b)(4).The olh (lot# 131278) was returned for investigation.Upon visual inspection, the blue plunger insert was missing, and the complaint was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCOMPASS SYNERGY CLAMP (LONG)
Common Device Name
ENCOMPASS SYNERGY CLAMP (LONG)
Manufacturer (Section D)
ATRICURE, INC.
7555 innovation way
mason OH 45040
Manufacturer (Section G)
ATRICURE, INC.
7555 innovation way
mason OH 45040
Manufacturer Contact
john ehlert
7555 innovation way
mason, OH 45040
5136448220
MDR Report Key17875912
Report Number3011706110-2023-00035
Device Sequence Number15227627
Product Code OCL
UDI-Device Identifier10840143905834
UDI-Public(01)10840143905834(17)260601(10)131278
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K210477
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2026
Device Model NumberOLH
Device Catalogue NumberA001143
Device Lot Number131278
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2023
Initial Report FDA Received Date10/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-