| Catalog Number |
BSTPULR6250SH |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
In this event it is reported that protaper ultimate shaper 25mm x6 broke during use.The broken part remains in the root canal.No injury.
|
| |
|
Additional Manufacturer Narrative
|
|
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: returned protaper ultimate shaper file 25mm is actually broken in the active part (fatigue).No material defect was found during analysis of the rupture pattern.No unused file is available for evaluation.Nothing unusual to report was found during dhr review (batch #1772181).Root causes are not identified.We will track this kind of event and monitor the trend.
|
| |
|