|
Mm 10.5.23 device report from canada reports an event as follows: it was reported that the instruments in question cannot be cleaned and disinfected in accordance with the mifu.On (b)(6) 2023 while inspecting the instruments with a borescope in mdrd, it was found that the cannulation in the product transitions from narrow to wide to narrow again, creating difficulties with manual cleaning.There was no reported patient or procedure interaction.No further information is available.This report is for a viper 2 system set screw inserter, self-retaining.This is report 2 of 8 for pc-(b)(4).
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|