• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC VIPER 2 SYSTEM SET SCREW INSERTER, SELF-RETAINING; DRIVER, PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC VIPER 2 SYSTEM SET SCREW INSERTER, SELF-RETAINING; DRIVER, PROSTHESIS Back to Search Results
Catalog Number 286735400
Medical Device Problem Codes Contamination /Decontamination Problem (2895); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2023
Type of Reportable Event Malfunction
Event or Problem Description
Device report from canada reports an event as follows: it was reported that the instruments in question cannot be cleaned and disinfected in accordance with the mifu.On (b)(6) 2023 while inspecting the instruments with a borescope in mdrd, it was found that the cannulation in the product transitions from narrow to wide to narrow again, creating difficulties with manual cleaning.There was no reported patient or procedure interaction.No further information is available.This report is for a viper 2 system set screw inserter, self-retaining.This is report 6 of 8 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER 2 SYSTEM SET SCREW INSERTER, SELF-RETAINING
Common Device Name
DRIVER, PROSTHESIS
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key17878509
Report Number1526439-2023-01940
Device Sequence Number7695199
Product Code HWR
UDI-Device Identifier10705034214580
UDI-Public(01)10705034214580
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number286735400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/19/2023
Initial Report FDA Received Date10/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-