• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA ¿MAYFIELD COMPOSITE SERIES BASE UNIT, STANDARD¿; ¿COMPOSITE SERIES¿

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA ¿MAYFIELD COMPOSITE SERIES BASE UNIT, STANDARD¿; ¿COMPOSITE SERIES¿ Back to Search Results
Catalog Number ¿A3101¿
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
This is 2 of 2 reports linked to mfg report number 3004608878-2023-00174: a facility reported that a surgeon observed an excess movement on a mayfield composite series base unit (a3101) during an unspecified surgical procedure.No information has been provided on patient consequence or surgical delay.
 
Additional Manufacturer Narrative
The mayfield composite series base unit, standard (a3101) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - investigation of the returned unit when set up could not duplicate any movement even when put under pressure.The handle tests are all within tolerances, and the unit passed all specific functional testing.Root cause - the complaint is not confirmed.Could not duplicate movement.The unit was recently repaired and does not display any type of movement.No further investigation is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
¿MAYFIELD COMPOSITE SERIES BASE UNIT, STANDARD¿
Common Device Name
¿COMPOSITE SERIES¿
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key17881320
Report Number3004608878-2023-00175
Device Sequence Number15393739
Product Code FWZ
UDI-Device Identifier10381780268161
UDI-Public10381780268161
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number¿A3101¿
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/12/2023
Initial Report FDA Received Date10/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/16/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1