Catalog Number 8065751013
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported ophthalmic aspiration tip was broken during surgery.The surgery was completed without any patient harm.
 
Additional Manufacturer Narrative
The product under investigation is not a serviceable device.Therefore, a service record review was not performed.However, the tip was returned for testing on this investigation.A non-conformance-based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.The tip was received for testing on this investigation.A visual assessment of the returned sample revealed it to be broken into two halves, confirming the reported event.However, although the reported event can be confirmed, how the tip became nonconforming could not be determined conclusively.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTREPID AUTOSERT IOL HANDPIECE, I/A TIP
Common Device Name
APPARATUS, CAUTERY, RADIOFREQUENCY, AC-POWERED
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17883257
Report Number2028159-2023-01363
Device Sequence Number15341461
Product Code HQC
UDI-Device Identifier00380657510139
UDI-Public00380657510139
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112425
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065751013
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer10/02/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/11/2023
Supplement Date Received by Manufacturer10/25/2023
Initial Report FDA Received Date10/05/2023
Supplement Report FDA Received Date11/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INFINITI VISION SYSTEM.