| Catalog Number |
MK31501 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
09/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an ongoing cpr the patient monitoring system failed to provide eco2 data which made clinical decisions in this situation more challenging.It was confirmed the resuscitation was successful that the patient was transferred to the icu.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation has just started; results will be provided in a follow-up report.H3 other text : on-going.
|
| |
|
Additional Manufacturer Narrative
|
|
T was reported the etco2 was functioning when undocked but once the m540 is docked, the etco2 reading disappears.A spare m540 was installed which replicated the reported issue then a spare m500 was installed.Once the m500 was put into place, the m540 was able to be docked while still displaying the etco2 readings.Upon draeger's receipt of the iacs, the system was evaluated where it was confirmed the cause of the issue was due to a faulty daughterboard.The daughterboard (pcb m500 daughter brd v7) was replaced in the m500 then sent back to the customer for use.
|
| |
|
Event or Problem Description
|
|
It was reported that during an ongoing cpr the patient monitoring system failed to provide eco2 data which made clinical decisions in this situation more challenging.It was confirmed the resuscitation was successful that the patient was transferred to the icu.
|
| |
|