|
Clinical adverse event received for pain.Event is considered mild.Unknown if event is considered serious.Event is possibly related to both device and procedure.Date of implant: (b)(6) 2012, date of event: (b)(6) 2023, (right knee).Treatment: ct scan for evaluation of interfaces.
|
|
Product complaint # (b)(4).D4-expiration date of the device was not provided.H6 component code: appropriate term/code not available (g07002) used to capture no findings available.Investigation summary: no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Device history batch: null.Device history review: null.
|