• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC CAVITRON 300 SERIES G310SCLR(DNA); SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC CAVITRON 300 SERIES G310SCLR(DNA); SCALER, ULTRASONIC Back to Search Results
Catalog Number G310SCLR
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event or Problem Description
While the customer was using a cavitron 300 series g310, they allege that the steri-mate and the inserts heated up and got hot.No injury was reported from the alleged event.
 
Additional Manufacturer Narrative
Investigation: ec2 main board upgrade.Could not replicate customer complaint(unit heating up) water flow is fine, ran unit for an hour without any problem.Battery icon will show low battery until food switch is activated this is.Normal.I upgraded main board to improved performance of the unit.Replaced damaged/worn components and recalibrated unit to factory specs.No water hose, power cord received for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAVITRON 300 SERIES G310SCLR(DNA)
Common Device Name
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
york, PA 17401
7178494593
MDR Report Key17885864
Report Number2424472-2023-00048
Device Sequence Number6976372
Product Code ELC
UDI-Device IdentifierD00382703121
UDI-PublicD00382703121
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K150535
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 11/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberG310SCLR
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/04/2023
Supplement Date Received by Manufacturer10/04/2023
Initial Report FDA Received Date10/06/2023
Supplement Report FDA Received Date11/13/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-