| Catalog Number |
G310SCLR |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
While the customer was using a cavitron 300 series g310, they allege that the steri-mate and the inserts heated up and got hot.No injury was reported from the alleged event.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: ec2 main board upgrade.Could not replicate customer complaint(unit heating up) water flow is fine, ran unit for an hour without any problem.Battery icon will show low battery until food switch is activated this is.Normal.I upgraded main board to improved performance of the unit.Replaced damaged/worn components and recalibrated unit to factory specs.No water hose, power cord received for evaluation.
|
| |
|