| Model Number |
02-01894 |
| Medical Device Problem Codes |
Moisture Damage (1405); Smoking (1585)
|
| Health Effect - Clinical Code |
Superficial (First Degree) Burn (2685)
|
| Date of Event |
09/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The patient reported that the monitor was damaged due to a smoking charging cord.The patient reacted and threw water on the smoking charging code.The monitor no longer functions due to the water and charging port is damaged.The patient reported that she received a small brun.The patient did not require medical attention.The devcie was returned.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that the monitor was damaged due to a smoking charging cord.The monitor no longer functioned.The monitor and the charging cord was returned for investigation.Engineering evaluation was able to confirm the reported complaint of "monitor damaged due to a smoking cord".The most likely root cause of the reported event is the due to an electrical fault within usb-a/usb-c charging cord.The component failure is further being investigated by philip's am&d.
|
| |
|
Additional Manufacturer Narrative
|
|
**udi related data quality updates only**.
|
| |
|