• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO MEDICAL CORPORATION DESTINATION PERIPHERAL GUIDING SHEATH; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO MEDICAL CORPORATION DESTINATION PERIPHERAL GUIDING SHEATH; INTRODUCER, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A1: patient identifier: requested, not provided.A2: age & date of birth: requested, not provided.A3: patient sex: requested, not provided.A4: weight: requested, not provided.A5: ethnicity: requested, not provided.A6: race: requested, not provided.D6a: implanted date: device was not implanted.D6b: explanted date: device was not explanted.E3: occupation: requested, unknown.A review of the device history record of the product code and lot number combination was conducted with no findings.The actual device was not available for product evaluation.The complaint cannot be confirmed for deformation of the sheath tip because the sample was not returned for assessment.The exact root cause could not be determined.Historically, deformations of the sheath tip occur during insertion, as the destination sheath tip is atraumatic and designed to prevent damage to the artery if faced with resistance.Review of the device history record (dhr) showed there were no issues noted during manufacture of the product that could have contributed to this complaint condition.Currently no action is recommended since this risk evaluation is within the predetermined limits in the design failure mode and effects analysis (dfmea).
 
Event or Problem Description
The user facility reported that upon opening the packaging of the destination product, it was noticed that the grey tip of the destination has lifted off or is coming off from the sheath.The procedure was completed successfully.There was no impact of event on patient.Additional information was received on 18sep2023: the device was not used on the patient.The patient was stable.There was no known affect as a result of the event.Additional information was received on 19sept2023: it was confirmed that the product was not used on the patient and there is no known effect on the patient as it was discovered upon receipt/preparation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DESTINATION PERIPHERAL GUIDING SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer (Section G)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer Contact
gina digioia
950 elkton blvd
elkton, MD 21921
6402040886
MDR Report Key17885959
Report Number1118880-2023-00410
Device Sequence Number7551496
Product Code DYB
UDI-Device Identifier00389701008912
UDI-Public00389701008912
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K172995
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/01/2025
Device Model NumberN/A
Device Catalogue Number54-54501
Device Lot Number0000375073
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2023
Initial Report FDA Received Date10/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-