The subject device was returned to olympus for evaluation.During inspection and testing, the allegation was confirmed.The knife wire had broken.The broken portion was scorched and melted.The outer diameter of the knife wire was measured, and no abnormalities were identified.The length of the coated portion and the knife wire presented no defects.There were no abnormalities identified that could have led to the breakage of the knife wire.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
|
An olympus employee reported on behalf of a customer, during a therapeutic papillotomy of the duodenum, the knife wire tip broke when the electricity was turned on for a second time.The procedure was completed using another similar device.There was no report of patient harm associated with this event.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation the specific cause of phenomenon could not be identified.However, multiple factors may have contributed to the knife wire breaking off: 1.The device was not protruded enough from the endoscope until the rear end of the cutting wire was in the field of view.2.Due to the situation of ¿1¿ description, the cutting wire and the endoscope were close to each other.3.The output was activated in state of ¿2¿ description.This might have led to an electrical discharge between the cutting wire and the distal end of the endoscope.4.An electrical discharge possibly occurred, and the cutting wire became hot instantly.That might have caused the cutting wire to break.The occurrence of the reported problem can be prevented by adhering to the instructions for use (ifu) which states the following: ¿since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.When inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.Do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.If you feel the cutting is blunt, withdraw the device from the scope to examine if there is any peel off and tear at the coating portion.¿ olympus will continue to monitor field performance for this device.
|