| Model Number |
20432030 |
| Medical Device Problem Codes |
Temperature Problem (3022); Excessive Heating (4030)
|
| Health Effect - Clinical Code |
Burn(s) (1757)
|
| Date of Event |
09/20/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The item in question was returned to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a surgery, the heating element (optitherm 20432030) burned the patient's skin.The heating element was connected to the thermoflator and probably overheated.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation of the product was completed on 2024-01-09.The optitherm heating element and the thermoflator were inspected.The thermoflator is working properly, no defect was found regarding the reported issue.The optitherm heating element was also checked.It was found that the lemo connector at the end had been opened by the customer and reassembled incorrectly (the pins were twisted after reassembly).The faulty behavior of the optitherm heater described by the customer could be reproduced if the lemo plug on the optitherm is plugged into the device with only one pin counterclockwise.The heater then heats continuously and becomes very hot after a short time, which can lead to skin burns.An inquiry to the lemo company also confirmed that the inside of the returned lemo plug is not an original part and that this can lead to incorrect assembly after opening.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Search Alerts/Recalls
|