| Model Number |
30001001001110013 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device powered on without display.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed and attributed to the led backlight module.The led backlight module was replaced to resolve the report.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report updates information based upon your request which differs from the initial medwatch report filed.Please reference sections d1 updated, d4 model # updated, d4 catalog # removed, d4 primary udi # updated.
|
| |
|
Search Alerts/Recalls
|