| Model Number |
H74939148301510 |
| Medical Device Problem Codes |
Break (1069); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/12/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that there was a shaft break and inability to remove a fragment of the device.The 100% stenosed target lesion was located in the moderately tortuous and severely calcified superficial femoral artery (sfa).A 3.0mm x 150mm x 150cm fg sterling balloon catheter was advanced for dilation.However, the balloon was fractured in the distal portion of the balloon that contained that distal ro marker about 5-10 mm from the tip.The majority of the catheter was removed but the distal tip was stuck in the mid sfa.Physician attempted to snare the device out numerous times but was unable to retrieve it.The physician then proceeded with angioplasty, followed by placing a stent, and yet the piece was not moving or obstructing flow.After seeing the post angiogram result, patient had significant improvement in flow and the physician left the fractured piece of the balloon catheter.The procedure was completed with another of same device.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter address 1: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being filed to correct the model number and unique identifier (udi) # in response to an fda request.E1: initial reporter address 1: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that there was a shaft break and inability to remove a fragment of the device.The 100% stenosed target lesion was located in the moderately tortuous and severely calcified superficial femoral artery (sfa).A 3.0mm x 150mm x 150cm fg sterling balloon catheter was advanced for dilation.However, the balloon was fractured in the distal portion of the balloon that contained that distal ro marker about 5-10 mm from the tip.The majority of the catheter was removed but the distal tip was stuck in the mid sfa.Physician attempted to snare the device out numerous times but was unable to retrieve it.The physician then proceeded with angioplasty, followed by placing a stent, and yet the piece was not moving or obstructing flow.After seeing the post angiogram result, patient had significant improvement in flow and the physician left the fractured piece of the balloon catheter.The procedure was completed with another of same device.There were no patient complications reported.
|
| |
|
Search Alerts/Recalls
|