| Catalog Number |
242049 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Stripper tip fractured during sampling.Patient's current condition: inaccessible foreign body left in place.
|
| |
|
Additional Manufacturer Narrative
|
|
Fh industrie procedures did not clearly define the rules for reporting incidents to fda when the devices were no longer being distributed into the us market.The vigilance officer at the time was unaware that these reporting rules had to be applied even to products that were no longer distributed in the united states and that were inactivated in the fda registration & listing database.The medwatch reports that were not filed are not connected to any recalls and had no impact on patient safety.A capa was opened to determine root cause and corrective action.
|
| |
|