U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INCEPTA CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INCEPTA CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) Back to Search Results
Model Number N160
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) and right ventricular (rv) defibrillation lead exhibited high out-of-range shock impedance measurements greater than 125 ohms for the last year and a half.Technical services (ts) reviewed troubleshooting options and possible causes, including possible lead calcification.This crt-d system remains in service, and will continue to be monitored at this time.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that there was no conclusive evidence that the observed out-of-range impedance measurements were due to a product performance issue or an inadequate lead-to-device connection.However, intermittent, out-of-range shock lead impedance measurements with no conclusive evidence of a performance issue or an inadequate lead-to-device connection are likely the result of the low-energy test signal utilized for daily automatic or in-clinic commanded shock lead impedance testing.A software update was released in 2015 to further improve consistency of the impedance test results and provide the clinician with additional diagnostic tools, including programmable shock lead impedance alert limits.
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) and right ventricular (rv) defibrillation lead exhibited high out-of-range shock impedance measurements greater than 125 ohms for the last year and a half.Technical services (ts) reviewed troubleshooting options and possible causes, including possible lead calcification.This crt-d system remains in service, and will continue to be monitored at this time.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCEPTA CRT-D
Common Device Name
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17890246
Report Number2124215-2023-55439
Device Sequence Number6994844
Product Code LWP
UDI-Device Identifier00802526480997
UDI-Public00802526480997
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P010012/S255
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/23/2016
Device Model NumberN160
Device Catalogue NumberN160
Device Lot Number967396
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/13/2023
Supplement Date Received by Manufacturer11/09/2023
Initial Report FDA Received Date10/06/2023
Supplement Report FDA Received Date12/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/06/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
-
-