| Catalog Number |
09180443001 |
| Medical Device Problem Code |
Application Program Problem (2880)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The software version is 3.03.13.The issue was resolved by manually importing the values into the infinity production server.The investigation determined that the event was caused by a software issue.The software issue was triggered by the end-of-day processing and caused the removal of the serum indices within the interference table because the import date of the cobas configuration was different from when it was exported.This issue has already been addressed.
|
| |
|
Event or Problem Description
|
|
The initial reporter alleged that comments that contain information needed for the interpretation of associated results are missing in the cobas infinity lab core license 3.X.The reporter stated that the missing iserum indices (si) values in the interference table prevented the addition of interpretation comments to the patient results.The values were reportedly present in the test server, but when the configuration was imported to the production server from the test, the values were missing.There was no report of any patient harmed.
|
| |
|