• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS INFINITY LAB CORE LICENSE 3.X; CALCULATOR/DATA PROCESSING MODULE FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS INFINITY LAB CORE LICENSE 3.X; CALCULATOR/DATA PROCESSING MODULE FOR CLINICAL USE Back to Search Results
Catalog Number 09180443001
Medical Device Problem Code Application Program Problem (2880)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/08/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The software version is 3.03.13.The issue was resolved by manually importing the values into the infinity production server.The investigation determined that the event was caused by a software issue.The software issue was triggered by the end-of-day processing and caused the removal of the serum indices within the interference table because the import date of the cobas configuration was different from when it was exported.This issue has already been addressed.
 
Event or Problem Description
The initial reporter alleged that comments that contain information needed for the interpretation of associated results are missing in the cobas infinity lab core license 3.X.The reporter stated that the missing iserum indices (si) values in the interference table prevented the addition of interpretation comments to the patient results.The values were reportedly present in the test server, but when the configuration was imported to the production server from the test, the values were missing.There was no report of any patient harmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS INFINITY LAB CORE LICENSE 3.X
Common Device Name
CALCULATOR/DATA PROCESSING MODULE FOR CLINICAL USE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17891787
Report Number1823260-2023-03224
Device Sequence Number7289700
Product Code JQP
UDI-Device Identifier07613336177129
UDI-Public07613336177129
Combination Product (Y/N)Y
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number09180443001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/08/2023
Initial Report FDA Received Date10/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-