| Catalog Number |
VA80104 |
| Medical Device Problem Codes |
Break (1069); Difficult to Insert (1316); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, photos were provided for review.The investigation of the reported event is currently underway.H10: (expiration date: 04/2026).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: device pending return.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to an angioplasty procedure, the pta balloon catheter allegedly had a break.It was further reported that the guidewire was allegedly difficult to be inserted into the balloon.The procedure was completed by using another device.There was no patient contact.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to an angioplasty procedure, the pta balloon catheter allegedly had a deformation.The procedure was completed by using another device.There was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: additional information was received, and the file was reassessed for reportability and determined to be no longer reportable.Since an initial mdr was submitted, therefore, the file will remain assessed as a malfunction.H10: d4 (expiration date: 04/2026), g3.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Device pending return.
|
| |
|
Search Alerts/Recalls
|