| Catalog Number |
07.01759.001 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a virage ball-tipped probe was missing the ball tip intra-operatively.The fluoro confirmed the tip was not in the patient.There was a 2-minute delay and no impact on the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: "device evaluation anticipated, but not yet begun (02)" no longer applies but cannot be deleted.Corrections in h3.Additional information in h6: component, investigation type, findings, and conclusions.Inspection the returned device matches the information in the complaint file and was examined.Visual inspection revealed that the ball tip of the device was fractured off.Dhr review the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Potential root cause a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.Device usage this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
|
| |
|
Event or Problem Description
|
|
It was reported that a virage ball-tipped probe was missing the ball tip intra-operatively.The fluoro confirmed the tip was not in the patient.There was a 2-minute delay and no impact on the patient.
|
| |
|