Model Number OER-MINI
Medical Device Problem Code Device Reprocessing Problem (1091)
Health Effect - Clinical Codes Unspecified Infection (1930); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2023
Type of Reportable Event Serious Injury
Event or Problem Description
A customer reported to olympus that the unit (oer mini endoscope reprocessor) failed the test strip for the disinfectant.The customer initially asked for instructions on how to change the disinfectant and was provided the instruction manual for the device.At that time, the customer reported that there was an infection.No other additional information was reported.The customer was contacted multiple times, but no further information was received.
 
Event or Problem Description
It was additionally reported by the customer that the issue was resolved the same day it was reported.The facility ran another cycle and the test strips passed.No patients were affected.It was further clarified that no scopes were processed until the matter was fixed, and there were no infections/concern for infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Common Device Name
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
MDR Report Key17896477
Report Number2429304-2023-00321
Device Sequence Number7006811
Product Code FEB
UDI-Device Identifier04953170331619
UDI-Public04953170331619
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/15/2023
Report Date (Section F) 09/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOER-MINI
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date09/18/2023
Device Age4 YR
Event Location Hospital
Date Report to Manufacturer09/18/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/09/2023
Supplement Report FDA Received Date11/15/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;