W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number CXA230005
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 09/20/2023
Type of Reportable Event Death
Additional Manufacturer Narrative
H6: code c19 - a review of the manufacturing records indicated the lot met all pre-release specifications.H6: code b20 -the device remains implanted and was therefore not available for engineering evaluation by gore.According to the gore® excluder® conformable aaa endoprosthesis instructions for use (ifu), potential adverse events that may occur and/or require intervention include, but are not limited to bleeding and death.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
The following was reported to gore: on (b)(6) 2023, the patient underwent an endovascular procedure in the ascending aorta to treat a pseudoaneurysm utilizing a gore® excluder® conformable aaa endoprosthesis (aortic extender).Reportedly, the ascending aorta was curved and the device did not take the curve (proximal) and had not apposed the wall on the inner curve which led to a slight bird beak.Physician believed the device covered the defect (pseudoaneurysm) and it was a difficult application/angle to treat.The slight bird beak did not seem to affect the outcome of the procedure and patient is doing well after 2 days of post op.Additional information received from fsa on 09/28/2023: last night ((b)(6) 2023).Fsa was informed by a dr.(b)(6) (who was apparently also at the initial procedure) that the patient passed away on an unknown date after the procedure.The cause of death was persistent bleeding.No further information or details are available as to what led to the persistent bleeding.It was also reported that the aortic extender conformable was initially placed through the patient's neck and into the aortic arch near the heart.The patient also had a previous heart transplant.
 
Additional Manufacturer Narrative
Additional investigation determined no product problem or deficiency caused or contributed to an adverse event/incident for the patient; the initial medwatch and any supplemental report submitted under manufacturer report number 3007284313-2023-02811 was submitted in error and is hereby retracted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
jaskaran parhar
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key17900860
Report Number3007284313-2023-02811
Device Sequence Number3367090
Product Code MIH
UDI-Device Identifier00733132650934
UDI-Public00733132650934
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P200030
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Catalogue NumberCXA230005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2023
Supplement Date Received by Manufacturer11/30/2023
Initial Report FDA Received Date10/09/2023
Supplement Report FDA Received Date11/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age5 YR
Patient SexUnknown
-
-