| Model Number |
15 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device will not complete boot-up cycle during power-up in this state, the device may not be able to deliver defibrillation therapy if needed.There was no patient involvement reported during the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and was able to verify and duplicate the reported issue.The cause of the reported issue was isolated to a loose connection in the backlight pcb.Stryker secured the connection to resolve the issue.The device passed functional and performance testing and was returned to the customer for use.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device will not complete boot-up cycle during power-up.In this state, the device may not be able to deliver defibrillation therapy if needed.There was no patient involvement reported during the event.
|
| |
|