• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET GMBH MEERA ST MOBILE OPERATING TABLE; TABLE, OPERATING-ROOM, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET GMBH MEERA ST MOBILE OPERATING TABLE; TABLE, OPERATING-ROOM, AC-POWERED Back to Search Results
Model Number 710001B0
Medical Device Problem Codes Use of Device Problem (1670); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.E1b event site name: (b)(6).H3 other text: device not returned to manufacturer.
 
Event or Problem Description
On 2nd october 2023 getinge became aware of an issue with one of our mobile tables - 710001b0 - meera st eu without auto drive.As it was stated, the following course of events occurred.The doctor was on the right side of the device.The table was in the trendelenburg position titled to the right.After repositioning the table into zero position by performing the downward movement, the doctor's foot was pressed laterally by the base of the table leading to skin injury of the user.According to the provided information, the incident did not result in any major injury.Additional details of the event and user's outcome were requested, however, no information has been received to date.It was confirmed that the patient was not harmed in this event.We decided to report the issue based on the potential for serious injury if the situation, namely the user's foot being pressed by the base of the table, was to reoccur.
 
Additional Manufacturer Narrative
Getinge became aware of an issue with one of our mobile tables - 710001b0 - meera st eu without auto drive.As it was stated, the following course of events occurred.The doctor was on the right side of the device.The table was in the trendelenburg position titled to the right.After repositioning the table into zero position by performing the downward movement, the doctor's foot was pressed laterally by the column of the table leading to skin injury of the user.According to the provided information, the incident did not result in any major injury.It was confirmed that the surgery was not delayed and the patient was not harmed in this event.We decided to report the issue based on the potential for serious injury if the situation, namely the user's foot being pressed by the table¿s column, was to reoccur.With the investigation performed it was concluded that upon the event occurrence, the device was being used for the patient¿s treatment, and thus were also directly involved with the reported incident.As no malfunction connected to the reportable event occurrence was found, it was considered that the getinge device was up to its specification.A review of the received customer product complaints revealed that in the past there was no serious injury to the patient or user when this particular issue occurred.Comparing the number of complained devices to the number of investigated meera operating tables on the market we can conclude the failure ratio is (b)(4) for the issue investigated herein.The issue investigated herein is a single and isolated case.According to the customer, the doctor was on the right side of the device.The technician has stated that this operating table model does not perform any movement in the base that makes the occursion claimed by the customer possible, as the locking system has a mechanical lock located on the base on the head side.The ssu concluded that it is probable by the time of the event occurrence was with his foot on the top of the operating table¿s base and leaning against the column.When moving it down, one of the covers pressed the side of the doctor¿s foot, as it did not cause any major injury.In the instruction for use (ifu 7100.01 rev 2, page 25), the user is warned that when adjusting, moving or storing the or table, the staff, the patient and the accessories are exposed to pinching and shearing hazards, particularly in the area around the joints at the head rest, back and leg plates.The user shall always ensure that no one can be subjected to pinching or shearing action or injured in any other way and that the accessories do not collide with any nearby objects.In the ifu (ifu 7100.01 rev 2, page 27), the user is also warned that any objects left on the base of the or table will damage the casing as it moves.The user shall not put any objects on the or table base.It is likely that the user utilized the operating table disregarding safety notes and suggestions from the user manual.In summary and as a result of the performed root cause evaluation, it was concluded that based on available information it can be suspected that the most probable root cause of the reported issue, namely the user's foot being pressed by the table¿s column, is related to the user error.We currently do not have any information that would warrant further action towards the device manufacturing or devices on the market, however as per our complaint handling processes will continue to monitor the customer experiences with the device for any future information.E1h event site postal code: (b)(6).The correction of h3a device evaluated by manufacturer, h3b device not eval provide code, h3c if other provide code - explain fields deems required.This is based on the internal evaluation.Previous b1 adverse event / product problem: adverse event & product problem corrected b1 adverse event / product problem: adverse event.Previous d1 brand name: meera st eu without auto drive corrected d1 brand name: meera st mobile operating table.Previous d4 catalog #: 710001b0.Corrected d4 catalog #: n/a.Previous b5 describe event or problem: on 2nd october 2023 getinge became aware of an issue with one of our mobile tables - 710001b0 - meera st eu without auto drive.As it was stated, the following course of events occurred.The doctor was on the right side of the device.The table was in the trendelenburg position titled to the right.After repositioning the table into zero position by performing the downward movement, the doctor's foot was pressed laterally by the base of the table leading to skin injury of the user.According to the provided information, the incident did not result in any major injury.Additional details of the event and user's outcome were requested, however, no information has been received to date.It was confirmed that the patient was not harmed in this event.We decided to report the issue based on the potential for serious injury if the situation, namely the user's foot being pressed by the base of the table, was to reoccur.Corrected b5 describe event or problem: on 2nd october 2023 getinge became aware of an issue with one of our mobile tables - 710001b0 - meera st eu without auto drive.As it was stated, the following course of events occurred.The doctor was on the right side of the device.The table was in the trendelenburg position titled to the right.After repositioning the table into zero position by performing the downward movement, the doctor's foot was pressed laterally by the column of the table leading to skin injury of the user.According to the provided information, the incident did not result in any major injury.It was confirmed that the surgery was not delayed and the patient was not harmed in this event.We decided to report the issue based on the potential for serious injury if the situation, namely the user's foot being pressed by the table¿s column, was to reoccur.Previous h3a device evaluated by manufacturer: no.Corrected h3a device evaluated by manufacturer: yes.Previous h3b device not eval provide code: other.Corrected h3b device not eval provide code: n/a.Previous h3c if other provide code - explain: device not returned to manufacturer.Corrected h3c if other provide code - explain: n/a.Previous h6 medical device ¿ problem code: insufficient information|||3190.Corrected h6 medical device ¿ problem code: use of device problem|||1670.
 
Event or Problem Description
On 2nd october 2023 getinge became aware of an issue with one of our mobile tables - 710001b0 - meera st eu without auto drive.As it was stated, the following course of events occurred.The doctor was on the right side of the device.The table was in the trendelenburg position titled to the right.After repositioning the table into zero position by performing the downward movement, the doctor's foot was pressed laterally by the column of the table leading to skin injury of the user.According to the provided information, the incident did not result in any major injury.It was confirmed that the surgery was not delayed and the patient was not harmed in this event.We decided to report the issue based on the potential for serious injury if the situation, namely the user's foot being pressed by the table¿s column, was to reoccur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEERA ST MOBILE OPERATING TABLE
Common Device Name
TABLE, OPERATING-ROOM, AC-POWERED
Manufacturer (Section D)
MAQUET GMBH
kehler strasse 31
rastatt
GM 
Manufacturer (Section G)
MAQUET GMBH
kehler strasse 31
rastatt
GM  
Manufacturer Contact
holger ullrich
kehler strasse 31
rastatt 
GM  
MDR Report Key17903383
Report Number8010652-2023-00098
Device Sequence Number15428191
Product Code FQO
UDI-Device Identifier04046768136298
UDI-Public(01)04046768136298(11)191106
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 10/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number710001B0
Device Catalogue Number710001B0
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2023
Supplement Date Received by Manufacturer09/06/2024
Initial Report FDA Received Date10/10/2023
Supplement Report FDA Received Date09/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/06/2019
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-