| Model Number |
MI1000 MED-EL CONCERT PIN |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During in situ testing, affected channels were seen.Re-implantation is considered but has not been scheduled yet.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the currently available information, damage to the active electrode as might be caused by an external mechanical impact appears likely.However, to determine an exact root cause a device investigation of the explanted device is necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Event or Problem Description
|
|
During in situ testing, affected channels were seen.Re-implantation is considered but has not been scheduled yet.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the currently available information, damage to the active electrode as might be caused by an external mechanical impact appears likely.However, to determine an exact root cause a device investigation of the explanted device is necessary.The concerned device was explanted but has not been received for investigation yet.
|
| |
|
Event or Problem Description
|
|
During in situ testing, affected channels were seen.The user has been re-implanted.
|
| |
|
Event or Problem Description
|
|
During in situ testing, affected channels were seen.The user has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|