Catalog Number 04.01.0111
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 09/18/2023
Type of Reportable Event Serious Injury
Event or Problem Description
At about 1 month after the primary, the patient came in reporting instability and the cause of the instability is unknown.The surgeon revised the metaphysis, liner, and glenosphere.The surgery was completed successfully.
 
Additional Manufacturer Narrative
Batch review performed on 19 september 2023: lot 2304475: (b)(4) items manufactured and released on 10-july-2023.Expiration date: 2028-06-20.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event during the period of review.Additional device involved: batch review performed on 19 september 2023: reverse shoulder system 04.01.0207 lat.Glenosphere 36xø24.5 (k193175) lot 2243584a: (b)(4) items manufactured and released on 10-may-2023.Expiration date: 2028-04-18.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event during the period of review.Clinical evaluation performed by medacta medical affairs manager: revision 1 month after the primary rsa, due to instability reported by the patient.With the information at hand it is not possible to determine the cause of this instability.However, it may be possible that the cause could be insufficient re-establishment of soft tissue tension after the operation.No further conclusion can be drawn with the elements at hand.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVERSE SHOULDER SYSTEM HUMERAL REVERSE HC LINER Ø36/+0MM
Common Device Name
SHOULDER REVERSE LINER
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key17904607
Report Number3005180920-2023-00775
Device Sequence Number7777231
Product Code PHX
UDI-Device Identifier07630040706261
UDI-Public07630040706261
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K170452
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number04.01.0111
Device Lot Number2244281
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2023
Initial Report FDA Received Date10/10/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-