| Catalog Number |
21-7355-24 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the line was leaking; split tubing under plastic mount in the unknown set.Patient involvement unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: b3.Date of event is unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: additional information is provided for d.10, h.2, h.3 and h.6.Two samples were received for evaluation.Visual inspection revealed the sample was received in a plastic bag; no damage or defects were detected.Functional testing was performed, and the reported issue was unable to be replicated.The root cause could not be determined.Device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.For all enquiries or follow-up questions related to the record, do not use (b)(6).
|
| |
|