| Catalog Number |
21-7361-24 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Other text: b3.Date of event is unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the line was leaking; split tubing under plastic mount in the unknown set.Patient involvement unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
One sample was received for evaluation.Visual inspection revealed the sample was received in unused conditions in a plastic bag; no damage or defects were noted.Functional testing was performed, and the reported issue was able to be replicated.The root cause was unable to be determined.Device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.For all enquiries or follow-up questions related to the record, do not use regulatory.Responses@smiths-medical.Com located in sections g.1., please direct those to the following: regulatory.Responses@icumed.Com.
|
| |
|
Search Alerts/Recalls
|