| Model Number |
3301 |
| Medical Device Problem Code |
Data Problem (3196)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device won't register data, but it turns on.Patient involvement is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: b3: date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: no product sample was received; therefore, visual and functional testing could not be performed.The most probable cause is a malfunctioning main board or spo2 probe/spo2 cable.The reported issue could not be confirmed.If the product is returned, the manufacturer will reopen this complaint for further investigation.A device history report (dhr) review could not be completed as the dhr is housed at the supplier.
|
| |
|
Event or Problem Description
|
|
Additional information received: there is no patient involvement and no patient or clinical injury.The outcome was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
**udi related data quality updates only**.
|
| |
|