| Model Number |
3301A3 |
| Medical Device Problem Code |
Failure to Sense (1559)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Other, other text: b3: date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the device takes a long time to register on oxygen (o2); "if it registers one at all".Patient involvement is unknown.
|
| |
|
Event or Problem Description
|
|
Additional information received via email.There was no patient involvement and no patient or clinical injury.The outcome was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
Health impact, and evaluation codes: updated.No product was returned, and no photographic evidence was provided to aid in the investigation.As a result, a complaint investigation, product evaluation and problem confirmation cannot be performed.The exact cause could not be determined.However, the most probable cause was a faulty main board or spo2 probe/spo2 cable.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.If the product is returned the manufacturer will re-open the complaint for further device analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
D4 was corrected.Udi related data quality updates only.
|
| |
|