MAUDE Adverse Event Report: EXACTECH, INC. TRULIANT TIB IMP PSC INSERT SZ 4, 12MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, CEMENTED, POLYMER/METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. TRULIANT TIB IMP PSC INSERT SZ 4, 12MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, CEMENTED, POLYMER/METAL/POLYMER Back to Search Results
Catalog Number 02-022-44-4012
Medical Device Problem Codes Unstable (1667); Naturally Worn (2988); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Pain (1994); Scar Tissue (2060); Synovitis (2094); Joint Laxity (4526); Insufficient Information (4580); Joint effusion (4849)
Date of Event 08/15/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported via legal documentation the patient had a left knee replacement on (b)(6) 2021.Approximately 1 year after the initial procedure the patient had a left knee revision on (b)(6) 2022.There is no other patient demographic or medical history available.There is no information on the surgical procedure or patient outcome.There is no device return.There are no photos or other images of the device provided.No additional information is available.The patient has filed a short-form complaint in a coordinated action in alachua county with master case (b)(4).The consolidated long form complaint that applies to cases filed in this coordinated action alleges that patients filing suits in this coordinated action were required ¿to undergo revision surgeries due to severe, pain, swelling, and instability¿ due to ¿wear of the polyethylene components and resulting component loosening and/or other failure failures causing serious complications including tissue damage, osteolysis, permanent bone loss, and other injuries.¿ because the patient has filed a short-form complaint in this coordinated action, the patient appears to allege that the patient was injured as a result of wear of an exactech polyethylene device.
 
Additional Manufacturer Narrative
D10: concomitants: 5655180 02-020-11-0240 - truliant ps cem fem ps cem left sz 4.6692020 02-012-60-1440 - tru stem ext 14mm x 40mm.6839836 204-70-00 - tibial stem ext.Screw.6937334 02-022-45-4035 - truliant tray, cem sz 4f/3.5t.Pending investigation.
 
Additional Manufacturer Narrative
The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and instability or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.H6: corrected health effect, medical device, component, and investigation clinical codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT TIB IMP PSC INSERT SZ 4, 12MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, CEMENTED, POLYMER/METAL/POLYMER
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
gainesville 32653
3523771140
MDR Report Key17906248
Report Number1038671-2023-02497
Device Sequence Number12671344
Product Code JWH
UDI-Device Identifier10885862305299
UDI-Public10885862305299
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K171045
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 02/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number02-022-44-4012
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2023
Supplement Date Received by Manufacturer02/21/2025
Initial Report FDA Received Date10/10/2023
Supplement Report FDA Received Date02/21/2025
Date Device Manufactured02/26/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0023-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-