| Model Number |
20E |
| Medical Device Problem Code |
Sparking (2595)
|
| Health Effect - Clinical Codes |
Atrial Flutter (1730); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker contacted the customer to request additional information on the patient.The customer provided stryker with the available patient information.Patient fields in which information is not provided were intentionally left blank. stryker continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that during intervention on a patient their device sparked and cause the serious injury to the patient.The two thermal burns were found on the patient's chest after intervention.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that during intervention on a patient their device sparked and cause the serious injury to the patient.The two thermal burns were found on the patient's chest after intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Section dh6 clinical signs code grid of initial mdr indicates: atrial flutter.Section dh6 clinical signs code grid of initial mdr should indicate: no clinical signs, symptoms or conditions.The customer's device and electrodes are not returned to stryker for the evaluation.A cause of the reported issue could not be determined.H3 other text : device not returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: section d4 gtin of the initial medwatch report indicates: blank.Section d4 gtin of the initial medwatch report should have indicated: (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that during intervention on a patient their device sparked and cause the serious injury to the patient.The two thermal burns were found on the patient's chest after intervention.
|
| |
|