• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOFRADIM PRODUCTION SAS DEXTILE; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOFRADIM PRODUCTION SAS DEXTILE; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number DXT1510AR
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hematoma (1884); Pain (1994); Seroma (2069); Hernia (2240)
Date of Event 07/23/2023
Type of Reportable Event Serious Injury
Event or Problem Description
According to the literature source of a study conducted between 2019 and 2022, a retrospective study evaluated outcomes in patients who underwent the standardized total extra-peritoneal technique.In some patients, a spacemaker dissection balloon was used to dissect the extra-peritoneal space.Dextile anatomical mesh (10x15) or a competitor mesh was placed without fixation to cover all potential hernia defects.There were 416 patients in the dextile group and 540 patients in the competitor group.Postoperative complications related to the mesh included seroma, hematoma, pain, and hernia recurrence.Complications related to spacemaker included wound infection and delayed wound healing.Recurrences were due to either a lipoma or a true recurrence, requiring either surgical repair or conservative treatment.Pain was managed with analgesics, lidocaine injections, or pulsed radiofrequency ablation.Antibiotics were necessary for infection, and prolonged hospitalization was reported.
 
Additional Manufacturer Narrative
D10 concomitant products: dxt1510al, dextile dxt1510al 15x10cm left x1 (lot#unknown) unk spacemaker, unknown spacemaker balloon (lot#unknown) reference: faessen, 2023, the first experience with the dextile anatomical mesh in laparoscopic inguinal hernia repair.Hernia.10.1007/s10029-023-02855-3.Accepted: 27 july 2023.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXTILE
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
Manufacturer (Section G)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR   01600
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key17906591
Report Number9615742-2023-01363
Device Sequence Number12718394
Product Code FTL
UDI-Device Identifier10884521726512
UDI-Public10884521726512
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K192443
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Literature,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDXT1510AR
Device Catalogue NumberDXT1510AR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2023
Initial Report FDA Received Date10/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-