| Catalog Number |
283869 |
| Medical Device Problem Codes |
Break (1069); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This is report 1 of 2 for (b)(4).It was reported by the sales rep in switzerland that during an unknown procedure on an unknown date, it was observed that 5.0 mm barrel burr plus device was blocked while in use with an unknown handpiece device.Changed to another devices but had the same issue.It was further reported that fragments were generated.It was unknown how the procedure was completed with a delay of five minutes.There were no adverse patient consequences reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi:(b)(4).The lot number was unknown.D4, g1, h4: the lot number was unknown; therefore, the expiration date, manufacturing site name and device manufacture date were unknown.E3: reporter is a j&j sales representative.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: investigation summary = > the complaint device was received for evaluation.Upon reviewing the device, no structural anomalies were found.The inner shaft shows rests of biological matter.No anomalies were found on the black connector.Given that lot number was not provided, the manufacturing record evaluation (mre) review cannot be performed.If the lot number becomes available, the mre review will be performed.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.Based on the investigation findings, this complaint can be confirmed for the burr blockage.A possible root cause for the blockage experienced by the user can be attributed to the procedural variables, such handling of the device or product interaction during procedure; higher speeds applied on the handpiece can increase the temperature of the device's shaft, lack of adequate irrigation and excessive side loading can contribute to a blocked suction hole, as per ifu: adequate irrigation to the tip of the blade or burr is required to cool the blade and prevent accumulation of excised material in the surgical site.Excessive side loading may result in blade wear and degradation, as well as clogging and/or seizing.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|
| |
|
Search Alerts/Recalls
|
|