| Model Number |
GIF-N180 |
| Medical Device Problem Code |
Microbial Contamination of Device (2303)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus.Prior to the device being evaluated, it was sent to an independent laboratory for culture testing.The device tested positive for < 1 colony forming units (cfus) of unspecified revivable microorganisms which, is conforming to standard.The customer has not provided the cleaning sterilization and disinfection (cds) processes performed at the user facility however, this information has been requested.During the device evaluation it was uncovered that the glue of the bending section cover had discoloration and was separated.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus, the evis exera ii gastrointestinal videoscope tested positive for unexpected contamination.The scope was sampled once.During the sampling, the scope tested positive for > 80 colony forming units (cfus) of escherichia coli.The issue was found at reprocessing during a routine culture of the scope.All channels were tested.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing, however, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The following is included in the device ifu: "an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." olympus will continue to monitor field performance for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer provided the cleaning sterilization and disinfection (cds) processes performed at the user facility.The customer did confirm that there were no deviations or deficiencies concerning reprocessing of the scope.Additionally, the customer confirmed that there were no suspected patient infections due to the facility findings.The customer confirmed that the facility does not delay the start of precleaning on the equipment after use.The customer did not provide the specific steps taken during the pre-cleaning and manual cleaning process.For automated endoscope reprocessor (aer) treatment, a s3 et wd440 machine is used with anios et detergent and soluscope p/ endo high (disinfectant).The scope is dried using towel/paper, then blown dried using filtered compressed air.The scope is stored in a plasma typhoon storage.The customer confirmed that all channels were connected with tubes when the endoscope was setting into the aer.The customer confirmed that the concentration and expiration date of the disinfectant was controlled.The customer confirmed that the water quality of the rinse water was controlled and filtered.The customer confirmed that the water filter was replaced periodically in accordance with the instructions for use (ifu).The investigation is ongoing.A final report will be submitted upon completion of the investigation.
|
| |
|