DEPUY ORTHOPAEDICS INC US DURALOC CONST LINER 50X28; DURALOC IMPLANT : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 124110527
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Foreign Body Reaction (1868)
Date of Event 09/28/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # : pc-(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that head liner exchanged was preformed uneventfully.A significant amount of metalosis was evident leading surgeon to believe the locking ring had been disassociated for some time.No patient harm occurred.It was unknown, if there was a surgical delay.Doi- unknown.Dor- (b)(6) 2023.Affected side- right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A complaint database search did find additional results but no related reports against the provided product code and lot number combination(s).Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURALOC CONST LINER 50X28
Common Device Name
DURALOC IMPLANT : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17908064
Report Number1818910-2023-20595
Device Sequence Number12652419
Product Code KWZ
UDI-Device Identifier10603295018643
UDI-Public10603295018643
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P960054
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2018
Device Catalogue Number124110527
Device Lot Number469857
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/29/2023
Supplement Date Received by Manufacturer10/25/2023
Initial Report FDA Received Date10/10/2023
Supplement Report FDA Received Date10/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/10/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICUL/EZE BALL 28 +1.5 GR.; DURALOC DYNAMIC LOCK RING 50MM.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age85 YR
Patient SexFemale
-
-